MRI Safety Information | Advanced Bionics

MRI Safety Information

For Advanced Bionics Cochlear Implants, United States

Testing has demonstrated that the Advanced Bionics HiRes implant family is MRI Conditional. Conditions vary by geography. A patient with the implant may be safely scanned with MRI only under very specific conditions. Scanning under different conditions may result in severe patient injury.

MRI Safety Information by Type of Implant

Type of Implant Model Number MRI Field Strength (T) Spatial Gradient Field (T/m) Max Head SAR (W/kg) Max Body SAR (W/kg)
Clarion 1.0* MMT-5100 L/R MRI is contraindicated None None None
Clarion 1.2* AB-5100 L/R MRI is contraindicated None None None
Clarion CII Bionic Ear*     AB-5100H MRI is contraindicated None None None
HiRes 90K*, **     CI-1400 1.5T 2.5 T/m ≤ 1 W/kg ≤ 1.7 W/kg
HiRes 90K Advantage*, **     CI-1500 1.5T 2.5 T/m ≤ 1 W/kg ≤ 1.7 W/kg
HiRes Ultra CI-1600

1.5T

3.47 T/m 13.90 T/m*** ≤ 3.2 W/kg ≤ 2 W/kg
  3.0T** 6.9 T/m ≤ 2.6 W/kg ≤ 2 W/kg
HiRes
Ultra 3D
CI-1601 1.5T 20 T/m ≤ 3.2 W/kg ≤ 2 W/kg
  3.0T 20 T/m ≤ 2.6 W/kg ≤ 2 W/kg

* These devices are no longer sold in the EU or North America
** For MRI the magnet has to be removed surgically
*** With magnet removed


Select the Implant name below
for further MRI information


HiRes™ Ultra 3D Cochlear Implant


The implant components of the HiRes™ Ultra 3D cochlear implant are MR conditional.

The external components of the HiResolution™ Bionics Ear System are MR Unsafe and must be removed before entering a room containing an MRI scanner.

The HiRes Ultra 3D cochlear implants have a specifically designed magnet assembly that allows safe MRI scanning, without angular restrictions of the head, and without surgical removal of the magnet assembly or a bandaging protocol.



Testing has demonstrated that the HiRes Ultra 3D cochlear implant is MR Conditional. Unilateral and bilateral recipients with this device with the multi-magnet assembly in place can be safely scanned in a horizontal closed bore quadrature coil MR system meeting the following conditions:   

MRI Field Strength

1.5T

3.0T

Maximum Spatial Field Gradient 20 T/m 20 T/m
RMS Gradient Field 34.4 T/s 34.4 T/s
Peak Slew Rate 200 T/m/s 200 T/m/s
Maximum Whole Body Averaged SAR 2.0 W/kg 2.0 W/kg
Maximum Head Averaged SAR 3.2 W/kg 2.6 W/kg

When tested under scan conditions defined above, the HiRes Ultra 3D cochlear implant produced a maximum temperature rise of <3˚C after 15 minutes of continuous scanning.

Note: During the scan, patients might perceive auditory sensations. Adequate counseling of the patient is advised prior to performing the MRI. The likelihood and intensity of the auditory sensations can be reduced by selecting sequences with a lower Specific Absorption Rate (SAR) and slower gradient slew rates.

In MRI testing, the measured range of device image artifact radius extending from the HiRes Ultra 3D cochlear implant for spin and gradient echo sequences and all imaging planes are as follows (refer to Tables A-B):


Table A: Artifact at 3.0T MRI Field Strength

Implanted

Multi-magnet
Assembly

Artifact Range

Unilaterally In Place 5.5 - 6.9 cm
Unilaterally Removed 1.4 - 4.2 cm
Bilaterally In Place 6.1 - 7.4 cm
Bilaterally Removed 1.9 - 6.9 cm

Table B: Artifact at 1.5T MRI Field Strength

Implanted

Multi-magnet
Assembly

Artifact Range

Unilaterally In Place 4.1 - 6.5 cm
Unilaterally Removed 2.4 - 3.2 cm
Bilaterally In Place 5.7 - 8.2 cm
Bilaterally Removed 3.4 - 4.1 cm

Note: For cases that would clinically benefit from reduced device artifact (for example, some head or neck scans), the internal multi-magnet assembly is surgically removed and possibly replaced with a Temporary Non-Magnetic Plug before the recipient undergoes an MRI procedure. The HiRes Ultra 3D cochlear implant can withstand 5 replacement cycles.

For additional information regarding the use of an MRI scanner with a HiRes Ultra 3D Cochlear Implant, please contact Advanced Bionics Technical Support at technicalservices@advancedbionics.com


HiRes™ Ultra Cochlear Implant

Device HiRes Ultra
Instructions For Use HiResolution Bionic Ear System: HiRes Ultra
1.5T magnet removed Scanning ok under certain conditions
1.5T magnet in place with MRI Antenna Coil Cover and bandaging protocol CI-7521 Scanning ok under certain conditions
3.0T magnet removed Scanning ok under certain conditions






For additional information regarding the use of an MRI scanner with a HiRes Ultra device, please contact Advanced Bionics Technical Support at technicalservices@advancedbionics.com or 1.877.454.5051


Auditory Technical Services

Advanced Bionics provides one-on-one technical support for professionals – including surgeons, audiologists, clinicians, educators, and therapists.


The form has been submitted successfully!

Thank you for contacting us to assist with your hearing solutions. Count on us—a community of hearing health professionals and recipients of cochlear implants—to help you Hear and Be Heard™. A representative from Advanced Bionics will be in touch soon!

If your request needs immediate attention or you have additional questions, please contact Hear@AdvancedBionics.com.
Toll Free Phone:  +1.866.844.HEAR (4327) US and Canada

We're here to help

Our team is comprised of caring professionals and dedicated volunteers who provide you and your loved ones with information, education, and support for choosing, getting, and living with cochlear implants.

 








AB Privacy Policy AB takes your privacy seriously. We will never sell or transfer your personal information to other companies for marketing purposes.


US-Form-1


The form has been submitted successfully!

Thank you for contacting us to assist with your hearing solutions. Count on us—a community of hearing health professionals and recipients of cochlear implants—to help you Hear and Be Heard™. A representative from Advanced Bionics will be in touch soon!

If your request needs immediate attention or you have additional questions, please contact Hear@AdvancedBionics.com.
Toll Free Phone:  +1.866.844.HEAR (4327) US and Canada

Contact Us








AB Privacy Policy AB takes your privacy seriously. We will never sell or transfer your personal information to other companies for marketing purposes.


US-Form-4


D000027430 RevD