MRI Safety Information | Advanced Bionics

MRI Safety Information

For Advanced Bionics Cochlear Implants, Singapore

Testing has demonstrated that the Advanced Bionics HiRes implant family is MRI Conditional. Conditions vary by geography. A patient with the implant may be safely scanned with MRI only under very specific conditions. Scanning under different conditions may result in severe patient injury. 


MRI Safety Information by Type of Implant

Type of Implant Model Number MRI Field Strength (T) Spatial Gradient Field (T/m) Max Head SAR (W/kg) Max Body SAR (W/kg)
Clarion 1.0* MMT-5100 L/R MRI is contraindicated None None None
Clarion 1.2* AB-5100 L/R MRI is contraindicated None None None
Clarion CII Bionic Ear*     AB-5100H MRI is contraindicated None None None
HiRes 90K*     CI-1400 1.5T 2.5 T/m ≤ 1 W/kg ≤ 1.7 W/kg
HiRes 90K Advantage     CI-1500 1.5T 2.5 T/m ≤ 1 W/kg ≤ 1.7 W/kg
HiRes Ultra CI-1600

1.5T

3.47 T/m 13.90 T/m*** ≤ 3.2 W/kg ≤ 2 W/kg
  3.0T** 6.9 T/m ≤ 2.6 W/kg ≤ 2 W/kg
HiRes
Ultra 3D
CI-1601 1.5T 20 T/m ≤ 3.2 W/kg ≤ 2 W/kg
  3.0T 20 T/m ≤ 2.6 W/kg ≤ 2 W/kg

* These devices are no longer sold in the EU or North America
** For MRI the magnet has to be removed surgically
*** With magnet removed

Select the Implant name below for further MRI information

HiRes™ Ultra 3D Cochlear Implant 


Device HiRes Ultra 3D
Instructions For Use
HiResolution Bionic Ear System: HiRes Ultra 3D
1.5T magnet in place
Scanning ok under certain conditions
3.0T magnet in place Scanning ok under certain conditions

3.0T with the magnet in place

The HiRes Ultra 3D cochlear implants have a specifically designed multi-magnet assembly that allows safe MRI scanning with the magnet in place, without angular restrictions of the head, surgical removal of the magnet or a bandaging protocol.



 

HiRes™ Ultra Cochlear Implant 


Device HiRes Ultra
Instructions For Use
HiResolution Bionic Ear System: HiRes Ultra 
1.5T magnet removed
Scanning ok under certain conditions
1.5T magnet in place with MRI Antenna Coil Cover and bandaging protocol CI-7521 Scanning ok under certain conditions
3.0T magnet removed Scanning ok under certain conditions

1.5T with the magnet removed




1.5T with the magnet in place




3.0T with the magnet removed





Magnet Removal/Replacement Using the HiRes Ultra Magnet Tool Kit, CI‑1418

These tools must be sterilized prior to use. See the ”Guide for Reprocessing HiRes Ultra Reusable Tools” provided in the kit. 

HiRes Ultra Magnet Tool Kit contains Two (2) Magnet Insertion Tools and Two (2) Magnet Pusher Tools


HiRes™ 90k Cochlear Implant Family

Device HiRes 90k and HiRes 90K Advantage    
Instructions For Use HiResolution Bionic Ear System: HiRes 90K and 90K Advantage
1.5T magnet removed Scanning ok under certain conditions
1.5T magnet in place with MRI Antenna Coil Cover and bandaging protocol CI-7521 Scanning ok under certain conditions



MRI Antenna Coil Cover

The MRI Antenna Coil Cover is intended to be used along with a bandaging protocol to allow the patient to undergo an MRI procedure with the magnet left in place. Instructions for use must be followed, including the recommended bandaging protocol.

Compression Dressing prior to MRI Procedure with HiRes 90K Users

This document will provide general guidance for applying appropriate compression dressing over the site of the implanted device (and magnet) in order to perform an MRI procedure.


Required Materials



Hurt-Free Wrap – Hospital Grade

  • Latex Free
  • 5 cm width
  • Johnson and Johnson (or equivalent)


Flexible Tape

  • Latex Free
  • 2 – 2.5 cm width
  • 3M – Nexcare (or equivalent)


MRI Antenna Coil Cover CI-7521 - Advanced Bionics



Patient Headpiece



Marking Pen



Coach Self-Adhering Elastic Bandage 3 inch (8 cm) width

  • Johnson and Johnson


Procedure


1

Place patient in sitting position to allow access to the implant site.

NOTE: If the patient is bilaterally implanted, prepare to locate both implant devices using the following steps. Both implant locations must have compression dressing placed over their magnets.

2

Secure MRI Antenna Coil Cover over implant magnet site. Place the patient headpiece (with cable removed) over the implant site. The magnets will hold the headpiece in place.


Patient headpiece magnetically attached

3

Cut a piece of the Coach bandage that is long enough to wrap around the head once..

4

Wrap this piece around the head, so that the bandage covers the patient headpiece. Make this wrap tight.


Patient headpiece under bandage

5

Outline the position of the headpiece on the bandage using a marker or pen.


Marked bandage showing outline of headpiece

6

Slip out the patient headpiece, but keep the bandage in place.


Removing headpiece with bandage in place

7

Slide the MRI Antenna Coil Cover under the bandage, lining it up with the marked outline of the headpiece.


Inserting MRI Antenna Coil Cover 

MRI Antenna Coin Cover under marked outline

8

Measure head size and bandage length needed for compression wrapping. Take the remaining Coach bandage roll, and wrap the bandage around the head once, without stretching.


Remaining Coach bandage roll wrapped once around head

9

Mark the location on the bandage that is one full wrap around the head. This is the head circumference.


Mark bandage one full wrap

Marked head circumference around head

10

Unwrap the piece of bandage with the marked head circumference, and place it on a flat surface.


Marked head circumference

11

Unroll the remaining bandage roll. Fold over the bandage start and crease at the marked head circumference.


12

Cut or tear the remaining bandage where it overlaps with the bandage start. The resulting bandage piece length is twice the head circumference.


13

Apply compression bandaging. Wrap this cut piece, this time very tightly, by stretching the marked line an additional half turn around the head.

NOTE: After 1.5 turns, the marked line should be on the opposite side of the head from where you started the wrap.


14

Continue wrapping the bandage at 150% extension for an additional 1.5 turns, resulting in 3 full turns total.

NOTE: After 3 turns, the bandage should end where you started the wrap (further is acceptable also).



After the MRI

Once the MRI procedure has been completed, gently remove the Coach Wrap, tape, and the MRI Antenna Coil Cover.

Discard all components including the MRI Antenna Coil Cover.

The form has been submitted successfully!

Thank you for contacting us to assist with your hearing solutions. Count on us—a community of hearing health professionals and recipients of cochlear implants—to help you Hear and Be Heard™. A representative from Advanced Bionics will be in touch soon!

If your request needs immediate attention or you have additional questions, please contact Hear@AdvancedBionics.com.
Toll Free Phone:  +1.866.844.HEAR (4327) US and Canada

We're here to help

Our team is comprised of caring professionals and dedicated volunteers who provide you and your loved ones with information, education, and support for choosing, getting, and living with cochlear implants.

 







AB Privacy Policy AB takes your privacy seriously. We will never sell or transfer your personal information to other companies for marketing purposes.



SG-Form-4
-iprof-content


Auditory Technical Services

Advanced Bionics provides one-on-one technical support for professionals – including surgeons, audiologists, clinicians, educators, and therapists.

D000030406 RevB